CareFusion Smoke Check Meter Especificações Página 11

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 24
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 10
9
Note: the BabyCO meter should only be connected to a computer
that is manufactured in accordance with EN60950 1992/1993 –
‘Safety of Information Technology Equipment including Electrical
Business Equipment’.
Calibration
Calibration will remain stable to within 2% over one month and
typically to within 10% over 6 months. CareFusion supplies
calibration gas (20ppm CO in air) and recommends that the unit be
recalibrated on a 6 monthly basis. See page 15 for calibration
accessories.
To carry out the calibration locate the calibration switch on the right-
hand side of the instrument as shown below.
Push the slide switch to the CO - PPM position and wait for the unit
to display zero.
Screw the control valve firmly onto the cylinder and connect the gas
supply as shown below:
Serial
port
Calibration
Switch
Vista de página 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 ... 23 24

Comentários a estes Manuais

Sem comentários